1, Calibrate oxygen index text apparatus
Common
used oxygen limiting index test apparatus uses an oxygen analyzer to directly measure the oxygen
concentration and then display data. In order to ensure the accurate results, every
time before testing, we have to calibrate it in accordance with the
requirements and the procedure is carried out using the standard gas.
2,
Use fume hood
There
will be great amount of toxicity gas because of the material burning which can
not only pollute the environment but also harm to human body, thus according to
the standard, the test should be carried out in a fume hood.
3,
Clean combustion residue
Oxygenindex test apparatus bottom base is filled with diameters of 3 ~ 5 mm glass
beads with the filling height of 80 ~ 100mm to fully mix gas.
4,
Sample requirements
There
are strict requirements on various aspects of sample. To ensure test results
accuracy, we should strictly ensure sample shape, size, orientation direction,
surface and the internal conditions.
Conclusion
Oxygen
index test can be used for general quality control as a way to provide combustion
characteristics of materials burning in the lab. But it cannot be used to
assess the actual fire hazard while it can only be evaluated as an element of
fire risks. Oxygen index test method has several advantages such as good
repeatability, high accuracy, good reproducibility, ease of testing and wide application
thus has become an effective means to evaluate the combustion level.