Sweating guarded hotplate testing result calculation
formula:
Thermal Resistance: Rct = (Tm-Ta) .A / (H-
Hc) - Rct0
Dampness: Ret = (Ps-Pa) .A / (H- He) - Ret0
If necessary, you can calculate other
parameters according to the following formula:
Insulation rate: Wr = (1-b / a)%
ASTM F 1868A Rcf = Rct-Rcpb
ASTM F 1868B Ref = Ret-Repb
Breathable index: im = 0.060 Rcf / Ref
Water vapor transmission rate: Wd = 1 / Ret.ΦTm
When Tm = 35 ℃ when, ΦTm =
0.627W.h / g.
Crow value: clo = Rct /0.155 = 6.451 Rct
Thermal conductivity: k = 10-5.d / Rct
Wherein d is the thickness of the material
(mm)
sweating guarded hotplate
Other related symbols and is defined as:
Rct resistance, m2.K / W
Ret dampness, m2.Pa / W
imt breathable index, dimensionless
Rct0 for the determination of the thermal
resistance Rct determined instrument constant, m2.K / W
Ret0 for the determination of moisture
resistance Ret determined instrument constant, m2.Pa / W
Rcpb bare-board thermal resistance, with
Rct0
Repb bare board dampness, with Ret0
Wd water vapor transmission rate, g / m2hPa
ΦTm test plate surface temperature of saturated steam latent heat Tm
when, Wh / g
A test board area, m2 sweating guardedhotplate formula
Ta climate chamber air temperature, ℃
Temperature Tm test plate, ℃
Ts hot grommet temperature, ℃
Pa water vapor pressure (in a climate
chamber when the temperature Ta), Pa
Pm saturated water vapor pressure (when the
surface temperature of the test plate when Tm), Pa
va test sample over the surface of the air
velocity, m / s
sv standard deviation of flow velocity va,
m / s
Relative humidity R.H.,%
H provided to the test panel heating power,
W
sweating guarded hotplate testing
△ Hc measurement of thermal resistance Rct heating correction amount
of power
△ He dampness Ret correction amount determination heated power
α
△ Hc calculation of the amount of slope correction
β
△ He corrected the calculation of the amount of slope
k Thermal conductivity, W / m.K
d material thickness, mm
Wr Insulation rate
plus a empty plate heat (J / Cm2.S)
b have the test sample plus heat (J /
Cm2.S)
clo CLO value, the thermal resistance of a
representation unit. At a temperature of 21 ° C, the air flow does not exceed
0.1m / s environmental conditions, meditation (whose basal metabolism of 58 W /
m2) feel comfortable when they wear clothing insulation value of 1 Crow (clo)
value.